We are actively engaged in offering a premium quality Medical Spirometer. In tandem with the latest medical standards, our offered spirometer is procured from our reliable sources of industry. The offered range provides fast and accurate spirometry testing. Available in customized options, we offer this spirometer to our esteemed clients after assuring that our offered range is the best in quality and performance.
Features:
User-friendly
Highly effective
Durability
Spiro meter Technical specifications:
Temperature sensor: semiconductor (0-45C)
Flow sensor: bi-directional digital turbine
Volume range: 10 L
Flow range: 16 L/s
Volume accuracy: 3% or 50 mL
Flow accuracy: 5% or 200 mL/s
Dynamic resistance at 12 L/s: <0.5 cmH2O/L/s
Display type: graphic LCD - FSTN
Display resolution: 128 64 pixel
Keyboard: membrane, 6 Keys
Connectivity: USB, Bluetooth, RS232, wireless internal modem (for Telemedicine applications)
Power Supply: 4 1.5 V, AAA battery
Dimension: 2.45.71.2 inch (6014530 mm
Weight: 6.4 Oz (180 gram) with batteries
Measured parameters:
FVC, FEV1, FEV1/FVC%, FEV3, FEV3/FVC%, FEV6, FEV1/FEV6%, PEF, FEF25%, FEF50%, FEF75%, FEF25-75%, FET, Vext, Lung Age, FIVC, FIV1, FIV1/FIVC%, PIF, VC, IVC, IC, ERV, FEV1/VC%, VT, VE, Rf, ti, te, ti/t-tot, VT/ti, MVV
Spirometer with SpO2 Opt Technical specifications:
SpO2 range: 0-99%
SpO2 accuracy: 2% between 70-99% SpO2
Pulse Rate range: 30-254 BPM
Pulse Rate accuracy: 2 BPM or 2%
Measured parameters:
Basic: SpO2 [Baseline, Min, Max, Mean],
Pulse Rate [Baseline, Min, Max, Mean], T90 [SpO2<90%], T89 [SpO2<89%], T88 [SpO2<88%], T5 [SpO2>5%], Index [12s], SpO2 Events, Pulse Rate Events [Bradycardia, Tachycardia]
6 Minute Walk Test: T2 [SpO22%],
T4 [SpO24%], Recording time, Time
[Rest, Walking, Recovery], Walked Distance, Predicted Distance [Min, Standard],
Desaturation Area/Distance,
Borg Dyspnea [Baseline, End, Change],
Borg Fatigue [Baseline, End, Change]
Sleep Test: Desaturation Events, Desaturation Index (ODI), Desaturation [Value, Duration, Nadir], SpO2 [Min Drop, Max Drop],
Pulse Variations, Pulse Rate Index,
NOD 89 [SpO2<89%; >5 min],
NOD 4 [SpO2 Baseline-4%; >5 min],
NOD 90 [SpO2<90%; Nadir <86%; >5 min]